TÜRKAK Pressure Calibration
Accredited calibration for manometers, vacuum gauges, transducers and transmitters.
Instruments that measure pressure need calibrating at regular intervals if they are to keep delivering precise and accurate results.
To achieve the most accurate measurement, our expert team works in our temperature- and humidity-controlled laboratory with precise, reliable reference instruments from world-leading manufacturers.
Application area
- Manometers
- Vacuum gauges
- Transducers
- Transmitters
What you receive
Your calibration certificate explicitly reports the measurement traceability, the accreditation details and the condition of the instrument, including any conditions found outside tolerance and any adjustments made during calibration.
One point to note: our accredited pressure scope is not a continuous range. It consists of two separate bands — -0.8 … -0.1 bar (vacuum) and 2 … 20 bar. If your instrument’s operating range falls outside these bands, please get in touch.
Accreditation No AB-0276-K · TS EN ISO/IEC 17025:2017 · Revision 26.08.2025 / 01
Our accreditation scope
The table below is reproduced from the annex to our TÜRKAK accreditation certificate.
The entries in this scope are marked “At the customer’s site · In the laboratory” on the certificate. Whether a particular instrument can be calibrated at your premises depends on its type and the measurement conditions; we assess this together when you make a request.
2 scope entries — scroll the table to see them all.
| Instrument calibrated | Measurement range | Conditions | Expanded measurement uncertainty (k=2) |
|---|---|---|---|
| Bağıl BasınçAnalog ManometreSayısal ManometreGöstergeli Basınç Ölçer | -0,8 bar ≤ p ≤ -0,1 bar | Pnömatik | 1,1 · 10⁻³ bar |
| Bağıl BasınçAnalog ManometreSayısal ManometreGöstergeli Basınç Ölçer | 2 bar < p ≤ 20 bar | Pnömatik | 2,5 · 10⁻³ bar |
Standards used: EURAMET/cg-17